PrincipleFat thickness measured by reflectance of ultrasound. ULTRA – FOM 300 |
|
Known standard deviations |
|||
|
Item |
Repeatability s(r) |
Reproducibility s(R) |
Reliability |
|
1 operator, 1 instrument |
5,48 |
- |
Very good |
|
1 operator, 2 instruments |
- |
5,38 |
Very good |
|
1 instrument, 3 operators |
|
5,04 |
Very good |
|
3 instruments, 3 operators |
|
5,05 |
Very good |
|
Known
possibilities for error |
||
|
Error |
Frequency |
Error contribution expressed as standard
deviation |
|
Dry surface |
Very low |
Not known |
|
Contact failure between instrument and skin |
Very low |
Not known |
|
Wrong measurement place |
Sometimes* |
Not known |
|
Dirty transducer |
Very low |
Not known |
It is of great importance for the daily control, that you are able to check the measurement point after the measurements have taken place. You should therefore only use instruments with a marking device.
It should be noted / documented which instrument that are used, and for each instrument the program version or EPROM number should be recorded to make it possible to track the used algorithm.
|
Daily
control |
Periodic
control |
|
Control of instrument |
Control of instrument |
|
Every day Every day Every day |
|
Control of data |
Control of data |
|
Every month |
Every month |